中文名稱:湖泊沉積物成分分析標準物質(zhì)
英文名稱:Lake Sediment訂貨以英文名稱為準。
產(chǎn)品編號:NIES-CRM-31CAS:/
品牌:標準物質(zhì)產(chǎn)地:日本
標準值:見證書分子式:/
規(guī)格 市場價(元) 折后價(元) 供貨周期 數(shù)量 加入購物車 20g 0.00 0.00 3-4周 加入購物車
-
-
產(chǎn)品詳細介紹
Certified values of NIES CRM No.31 in the whole material as supplied
The certified values and uncertainties are expressed as mass fraction.
Element Unit Certified value Uncertainty Analytical method Na % 0.882 0.063 AAS, ICP-OES, INAA, XRF Al % 9.17 0.86 AAS, ICP-OES, ICP-MS, INAA, XRF Ca % 1.25 0.12 ICP-OES, ICP-MS, XRF Ti % 0.442 0.044 ICP-OES, ICP-MS, XRF Fe % 5.38 0.55 ICP-OES, ICP-MS, INAA, XRF P mg/kg 925 60 AAS, ICP-OES, XRF V mg/kg 154 11 ICP-OES, ICP-MS, INAA Cr mg/kg 43.3 3.0 ICP-OES, ICP-MS, INAA, XRF Mn mg/kg 978 92 ICP-OES, ICP-MS, INAA, XRF Ni mg/kg 25.3 2.8 ICP-OES, ICP-MS Cu mg/kg 53.1 4.6 ICP-OES, ICP-MS, INAA, XRF Zn mg/kg 121 13 ICP-OES, ICP-MS, INAA, XRF Pb mg/kg 25.1 2.5 ICP-OES, ICP-MS AAS: atomic absorption spectroscopy
ICP-OES: inductively coupled plasma optical emission spectrometry
ICP-MS: inductively coupled plasma mass spectrometry
ID-ICP-MS: isotope dilution-ICP-MS
INAA: instrumental neutron activation analysis
XRF: X-ray fluorescence spectroscopyReference values of NIES CRM No.31 in the whole material as supplied
The referece values are expressed as mass fraction.
Element Unit Reference value Uncertainty Analytical method Mg % 0.836 0.068 AAS, ICP-OES, ICP-MS, INAA, XRF K % 0.991 0.134 AAS, ICP-OES, ICP-MS, INAA, XRF Sc mg/kg 19.1 3.4 ICP-OES, ICP-MS, INAA Co mg/kg 18.1 1.9 ICP-OES, ICP-MS, INAA As mg/kg 13.9 1.5 ICP-MS, INAA Sr mg/kg 125 14 ICP-OES, ICP-MS, XRF Ba mg/kg 338 49 ICP-OES, ICP-MS, INAA, XRF La mg/kg 20.4 2.2 ICP-OES, ICP-MS, INAA Cd mg/kg 0.342 0.043 ICP-MS, ID-ICP-MS Certified values of NIES CRM No.31 in the HCl/HNO3/HClO4 extract
The certified values and uncertainties are expressed as mass fraction.
Element Unit Certified value Uncertainty Analytical method V mg/kg 133 23 ICP-OES, ICP-MS Mn mg/kg 881 85 AAS, ICP-OES, ICP-MS Ni mg/kg 22.2 2.3 ICP-OES, ICP-MS Cu mg/kg 50.6 8.1 ICP-OES, ICP-MS Zn mg/kg 110 13 AAS, ICP-OES, ICP-MS Pb mg/kg 22.0 3.0 ICP-OES, ICP-MS AAS: atomic absorption spectroscopy
ICP-OES: inductively coupled plasma optical emission spectrometry
ICP-MS: inductively coupled plasma mass spectrometryReference values of NIES CRM No.31 in the HCl/HNO3/HClO4 extract
The referece values are expressed as mass fraction.
Element Unit Reference value Uncertainty Analytical method Cr mg/kg 33.7 6.1 ICP-OES, ICP-MS Cd mg/kg 0.285 0.067 ICP-OES, ICP-MS Characterization
The property values of the material were statistically determined based on chemical analyses by 13 organizations (including 20 laboratories) using a wide range of methods. A property value satisfying the following conditions was accepted as a certified value: 1) the relative standard deviation associated with the mean of the laboratory means was under 10 %; 2) the number of laboratories contributing to the mean of the laboratory means was at least ten; 3) the number of analytical methods contributing to the mean of the laboratory means was at least two. The uncertainty attached to the certified values is the expanded uncertainty using a coverage factor k=2, corresponding to the half-width of a confidence interval of approximately 95 %. A property value failing to satisfy one or two of the NIES criteria for certification but supplying valuable additional information about the material is given as a reference value. All certified and reference values were determined based on dry mass.
Description of the Material
The CRM is supplied as fine powder in an amber glass bottle. It is dark grayish-yellow in color.
Preparation of the CRM
The origin of this material is sediment collected at Lake Kasumigaura, Japan in February 1999. The raw sediment was dried in the sun at NIES and the clods that formed were crushed, sieved through a 2 mm sieve, and dried at 105 oC overnight. This dry material was then ground in an alumina ball mill, and passed first through a 212 μm sieve and then through a 75 μm sieve. The sieved material was homogenized with a V-blender to yield 12.2 kg of powder suitable for use as the CRM. The final material was placed in amber glass bottles (20 g in each bottle, 560 bottles), and sterilized by 60Co irradiation (20 kGy). All procedures complied with ISO GUIDE 34.
Homogeneity
Mass fractions of multi-elements, including those for which certified values are given, were determined by XRF in material taken from 10 bottles selected from the total 560 bottles by stratified random sampling. The between-bottle variation evaluated by a one-way analysis of variance (ANOVA) showed the relative standard deviations between bottles for the analytes to be less than 1 %. The material, therefore, is sufficiently homogeneous for its intended use as a reference material.
Instructions for Use
- Care should be taken to avoid contamination when opening the bottles. It is desirable to use up the contents as quickly as possible after opening.
- This CRM should be kept tightly closed in its original bottle and stored in a desiccator at room temperature (≤30 oC).
- Prior to weighing portions for analysis, the contents of the bottle should be shaken gently.
- It is recommended that a sample intake of 0.05 g is the minimum for convenient handling.
- Precautions should be taken to avoid inhalation of the material.
- This CRM should not be used for purposes other than research. When disposing of the material, local laws concerning processing and disposal of waste materials should be strictly adhered to.
- The mass fractions of elements in this CRM are reported on a dry mass basis. This CRM, as received, has less than 1 % water as measured in NIES by drying for 4 h at 105 oC. Correction to dry mass should be determined by drying a separate sub-sample.
Expiry Date of Certification
The expiry date for the certified values of this CRM is December 2024 assuming that the recommended storage conditions are adhered to. NIES will notify via its website if any changes in the contents are recognized within the term of validity.
Collaborating Laboratories in Analysis
The certified and reference values for this CRM were based on analytical values from the following 13 participating organizations:
National Institute for Environmental Studies; China Institute of Atomic Energy; Elemental Analysis, Inc.; Environmental Control Center Co., Ltd.; Environmental Research Center Co., Ltd.; IDEA Consultants, Inc.; Japan Environment Sanitation Center; Japan Food Research Laboratories; Kanagawa Industrial Technology Center; Murata Keisokuki Service Co., Ltd.; Naitoh Environmental Science Co., Ltd.; Nippon Steel & Sumikin Technology Co., Ltd.; Shimadzu Techno-Research Inc.Appendices
Appendix Fig.1
Schematic flow-chart of the HCl/HNO3/HClO4 extraction method according to “The Methods for Sediment surveillance” (in Japanese) of the Ministry of the Environment, Japan (public notice, August 2012).Appendix Table 1
Information values of Mass fractions of C, N, Si and S in NIES CRM No.31 Lake SedimentElement Unit Mass Fraction Analytical Method C % 3.22 EA N % 0.373 EA Si % 25.1 Gravimetry, XRF S % 0.391 ICP-OES EA: elemental analysis
XRF: X-ray fluorescence spectroscopy
ICP-OES: inductively coupled plasma optical emission spectrometry