- 您現(xiàn)在的位置:首頁 >> 產(chǎn)品中心 >> 標準物質(zhì) >> IAEA標準物質(zhì)IAEA標準物質(zhì)
-
-
中文名稱:蛤蜊/蚌/厚殼縱簾蛤
英文名稱:Clam (Gafrarium tumidum)訂貨以英文名稱為準。
產(chǎn)品編號:IAEA-461CAS:/
品牌:標準物質(zhì)產(chǎn)地:中國
標準值:見證書分子式:/
規(guī)格 市場價(元) 折后價(元) 供貨周期 數(shù)量 加入購物車 8g 0.00 0.00 3-4周 加入購物車
-
-
產(chǎn)品詳細介紹
IAEA-461蛤蜊/蚌/厚殼縱簾蛤Clam (Gafrarium tumidum)
Sixty kg of clams (Gafrarium tumidum) were collected in Noumea, New Caledonia. Clams were
opened, the soft-shell part was frozen and then freeze dried. The dry material was first ground and
sieved at 250 μm. A portion of about 5 kg was micronized to achieve a final powder size <50 μm. The
homogenization was performed by mixing the material in a stainless steel rotating homogenizer
Moritz ERM-BB124 (Moritz, Chatou, France) for 14 days at a temperature of 20 (±2)°C and relative
humidity of 50%. After checking for the homogeneity of sample material, aliquots of about 8 g were
packed into pre-cleaned brown borosilicate glass bottles with polyethylene caps and then sealed in
plastic bags. The sample material was labelled as IAEA-461.Analyte
Unit
Certified value(a)
Expanded uncertainty(b)
Cr
mg kg-1
27.0
2.4
As
mg kg-1
30.2
1.5
Ca
mg kg-1
9.27 x103
0.90 x103
Cd
mg kg-1
0.569
0.040
CH3Hg
mg kg-1 as Hg
62.3 x10-3
6.9 x10-3
Co
mg kg-1
4.78
0.60
Cu
mg kg-1
26.1
2.0
Fe
mg kg-1
2.60 x103
0.22 x103
Hg
mg kg-1
0.390
0.040
Mn
mg kg-1
333
20
Ni
mg kg-1
106
8
Pb
mg kg-1
25.8
1.7
Se
mg kg-1
2.78
0.31
V
mg kg-1
6.86
0.67
Zn
mg kg-1
156
11
(a) Information on calculation of certified values Certified values are calculated from the accepted data sets, each set being obtained by a different laboratory and/or a different method of determination following ISO Guide 35 ‘Reference materials – General and statistical principles for certification’ [1].
(b) Information on method used for uncertainty estimation The uncertainty is expressed as a combined uncertainty with a coverage factor k = 2, corresponding to a level of probability 95% , estimated in accordance with the JCGM 100:2008 ‘Evaluation of measurement data – Guide to the expression of uncertainty in measurement’ [2] and ISO Guide 35 [1].